| Shielding room front-end ground | Shielding mesh not installed | 13.57 | −15.0 |
| RF shielding mesh installed at window | 2.67 | −4.78 |
| Anti-insect mesh installed at window | 2.70 | −4.55 |
| Shielding room back-end ground | Shielding mesh not installed | 11.95 | −16.1 |
| RF shielding mesh installed at window | 3.41 | −3.38 |
| Anti-insect mesh installed at window | 2.95 | −4.29 |
| Shielding room mesh ground | Shielding mesh not installed | 11.03 | −16.3 |
| RF shielding mesh installed at window | 2.71 | −3.62 |
| Anti-insect mesh installed at window | 3.15 | −5.22 |
| Shielding room isolated | Shielding mesh not installed | 11.7 | −15.7 |
| RF shielding mesh installed at window | 2.69 | −4.99 |
| Anti-insect mesh installed at window | 3.38 | −4.54 |